Learn About The Law
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Current as of January 02, 2025 | Updated by Findlaw Staff
(I) The test procedures described in this appendix are for qualification of initial designs and major modifications of accepted designs. Included in (V) of this appendix are suggested formats that may be used in submitting test results to RUS.
(II) Sample Selection and Preparation. (1) All testing must be performed on lengths removed sequentially from the same 3 pair, 22 gauge jacketed wire. This wire must not have been exposed to temperatures in excess of 38°C since its initial cool down after sheathing. The lengths specified are minimum lengths and if desirable from a laboratory testing standpoint longer lengths may be used.
(a) Length A shall be 10 ±0.2 meters (33 ±0.5 feet) long and must be maintained at 23 ±3°C. One length is required.
(b) Length B shall be 12 ±0.2 meters (40 ±0.5 feet) long. Prepare the test sample by removing the inner and outer jacket, shield, and core wrap, if present, for a sufficient distance on both ends to allow the insulated conductors to be flared out. Remove sufficient conductor insulation so that appropriate electrical test connections can be made at both ends. Coil the specimen with a diameter of 15 to 20 times its sheath diameter. Three lengths are required.
(c) Length C shall be one meter (3 feet) long. Four lengths are required.
(d) Length D shall be 300 millimeters (1 foot) long. Four lengths are required.
(e) Length E shall be 600 millimeters (2 feet) long. Four lengths are required.
(f) Length F shall be 3 meters (10 feet) long and must be maintained at 23± 3°C for the duration of the test. Two lengths are required.
(2) Data Reference Temperature. Unless otherwise specified, all measurements shall be made at 23± 3°C.
(III) Environmental Tests—(1) Heat Aging Test—(a) Test Samples. Place one sample each of lengths B, C, D, and E in an oven or environmental chamber. The ends of sample B must exit from the chamber or oven for electrical tests. Securely seal the oven exit holes.
(b) Sequence of Tests. After conditioning the samples are to be subjected to the following tests:
(i) Water Immersion Test outlined in (III)(2) of this appendix;
(ii) Water Penetration Test outlined in (III)(3) of this appendix; .
(iii) Insulation Compression Test outlined in (III)(4) of this appendix; and
(iv) Jacket Slip Strength Test outlined in (III)(5) of this appendix.
(c) Initial Measurements. (i) For sample B, measure the open circuit capacitance and conductance for each pair at 1 and 150 kilohertz and the attenuation at 150 kilohertz after conditioning the sample at the data reference temperature for 24 hours. Calculate the average and standard deviation for the data of the 3 pairs on a per kilometer (per mile) basis.
(ii) The attenuation at 150 kilohertz may be calculated from open circuit admittance (Yoc) and short circuit impedance (Zsc) or may be obtained by direct measurement of attenuation.
(iii) Record on suggested formats attached in (V) of this appendix or on other easily readable formats.
(d) Heat Conditioning. (i) Immediately after completing the initial measurements, condition the sample for 14 days at a temperature of 65± 2°C.
(ii) At the end of this period note any exudation of filling compound. Measure and calculate the parameters given in (III)(1)(c) of this appendix. Record on suggested formats attached in (V) of this appendix or on other easily readable formats.
(iii) Cut away and discard a one meter (3 foot) section from each end of length B.
(e) Overall Electrical Deviation. (i) Calculate the percent change in all average parameters between the final parameters after conditioning with the initial parameters in (III)(1)(c) of this appendix.
(ii) The stability of the electrical parameters after completion of this test must be within the following prescribed limits:
(A) Capacitance. The average mutual capacitance must be within 5 percent of its original value;
(B) The change in average mutual capacitance must be less than 5 percent over the frequency range of 1 to 150 kilohertz;
(C) Conductance. The average mutual conductance must not exceed 2 micromhos/kilometer (3.3 micromhos/mile) at a frequency of 1 kilohertz; and
(D) Attenuation. The attenuation must not have increased by more than 5 percent over its original value.
(2) Water Immersion Electrical Test—(a) Test Sample Selection. The 10 meter (33 foot) section of length B must be tested.
(b) Test Sample Preparation. Prepare the sample by removing the inner and outer jacket, shield, and core wrap, if present, for a sufficient distance to allow one end to be accessed for test connections. Cut out a series of 2.5 millimeter by 13 millimeter (0.1 inch by 0.5 inch) rectangular slots along the test sample, at 300 millimeter (1 foot) intervals progressing successively 90 degrees around the circumference of the wire. Assure that the wire core is exposed at each slot by slitting the inner jacket and core wrap if present. Place the prepared sample in a dry vessel which when filled will maintain a one meter (3 foot) head of water over 6 meters (20 feet) of uncoiled wire. Extend and fasten the ends of the wire so they will be above the water line and the pairs are rigidly held for the duration of the test.
(c) Capacitance and Conductance Testing. Measure the initial values of mutual capacitance and conductance of all pairs in each wire at a frequency of 1 kilohertz before filling the vessel with water. Be sure the wire shield is grounded to the test equipment. Fill the vessel until there is a one meter (3 foot) head of water on the wires.
(i) Remeasure the mutual capacitance and conductance after the wires have been submerged for 24 hours and again after 30 days.
(ii) Record each sample separately on the suggested formats attached in (V) of this appendix or on other easily readable formats.
(d) Overall Electrical Deviation. (i) Calculate the percent change in all average parameters between the final parameters after conditioning with the initial parameters in (III)(2)(c) of this appendix.
(ii) The stability of the electrical parameters after of the test must be within the following prescribed limits:
(A) Capacitance. The average mutual capacitance must be within 5 percent of its original value; and
(B) Conductance. The average mutual conductance must not exceed 2 micromhos/kilometer (3.3 micromhos/mile) at a frequency of 1 kilohertz.
(3) Water Penetration Testing. (a) A watertight closure must be placed over the jacket of length C. The closure must not be placed over the jacket so tightly that the flow of water through preexisting voids or air spaces is restricted. The other end of the sample must remain open.
(b) Test per Option A or Option B. (i) Option A. Weigh the sample and closure prior to testing. Fill the closure with water and place under a continuous pressure of 10± 0.7 kilopascals (1.5± 0.1 pounds per square inch gauge) for one hour. Collect the water leakage from the end of the test sample during the test and weigh to the nearest 0.1 gram. Immediately after the one hour test, seal the ends of the wire with a thin layer of grease and remove all visible water from the closure, being careful not to remove water that penetrated into the core during the test. Reweigh the sample and determine the weight of water that penetrated into the core. The weight of water that penetrated into the core must not exceed 1 gram.
(ii) Option B. Fill the closure with a 0.2 gram sodium fluorscein per liter water solution and apply a continuous pressure of 10± 0.7 kilopascals (1.5± 0.1 pounds per square inch gauge) for one hour. Catch and weigh any water that leaks from the end of the wire during the one hour period. If no water leaks from the sample, carefully remove the water from the closure. Then carefully remove the outer jacket, shield, inner jacket and core wrap, if present, one at a time, examining with an ultraviolet light source for water penetration. After removal of the inner jacket and core wrap, if present, carefully dissect the core and examine for water penetration within the core. Where water penetration is observed, measure the penetration distance. The distance of water penetration into the core must not exceed 127 millimeters (5.0 inches).
(4) Insulation Compression Test. (a) Test Sample D. Remove inner and outer jacket, shield, and core wrap, if present, being careful not to damage the conductor insulation. Remove one pair from the core and carefully separate, wipe off core filler and straighten the insulated conductors. Retwist the two insulated conductors together under sufficient tension to form 10 evenly spaced 360 degree twists in a length of 100 millimeters (4 inches).
(b) Sample Testing. Center the mid 50 millimeters (2 inches) of the twisted pair between two smooth rigid parallel metal plates measuring 50 millimeters (2 inches) in length or diameter. Apply a 1.5 volt direct current potential between the conductors, using a light or buzzer to indicate electrical contact between the conductors. Apply a constant load of 67 newtons (15 pound-force) on the sample for one minute and monitor for evidence of contact between the conductors. Record results on suggested formats attached in (V) of this appendix or on other easily readable formats.
(5) Jacket Slip Strength Test—(a) Sample Selection. Test sample E from (III)(1)(a) of this appendix.
(b) Sample Preparation. Prepare test sample in accordance with the procedures specified in ASTM D 4565–90a.
(c) Sample Conditioning and Testing. Remove the sample from the tensile tester prior to testing and condition for one hour at 50± 2°C. Test immediately in accordance with the procedure specified in ASTM D 4565–90a. A minimum outer jacket slip strength of 67 newtons (15 pound-force) is required. Record the load attained.
(6) Humidity Exposure. (a) Repeat steps (III)(1)(a) through (III)(1)(c)(iii) of this appendix for separate set of samples B, C, D and E which have not been subjected to prior environmental conditioning.
(b) Immediately after completing the measurements, expose the test sample to 100 temperature cyclings. Relative humidity within the chamber must be maintained at 90± 2 percent. One cycle consists of beginning at a stabilized chamber and test sample temperature of 52± 1°C, increasing the temperature to 57± 1°C, allowing the chamber and test samples to stabilize at this level, then dropping the temperature back to 52± 1°C.
(c) Repeat steps (III)(1)(d)(ii) through (III)(5)(c) of this appendix.
(7) Temperature Cycling. (a) Repeat steps (III)(1)(a) through (III)(1)(c)(iii) of this appendix for separate set of samples B, C, D and E which have not been subjected to prior environmental conditioning.
(b) Immediately after completing the measurements, subject the test sample to 10 cycles of temperature between −40°C and +60°C. The test sample must be held at each temperature extreme for a minimum of 1 1/2 hours during each cycle of temperature. The air within the temperature cycling chamber must be circulated throughout the duration of the cycling.
(c) Repeat steps (III)(1)(d)(ii) through (III)(5)(c) of this appendix.
(IV) Control Sample—(1) Test Samples. A separate set of lengths for samples A, C, D, and E must have been maintained at 23± 3°C for at least 48 hours before the testing.
(2) Repeat steps (III)(2) through (III)(5)(c) of this appendix except use length A instead of length B.
(3) Surge Test. (a) One length of sample F must be used to measure the breakdown between conductors while the other length of F must be used to measure core to shield breakdown.
(b) The samples must be capable of withstanding, without damage, a single surge voltage of 20 kilovolts peak between conductors, and 35 kilovolts peak between conductors and the shield as hereinafter described. The surge voltage must be developed from a capacitor discharge through a forming resistor connected in parallel with the dielectric of the test sample. The surge generator constants must be such as to produce a surge of 1.5 x 40 microseconds wave shape.
(c) The shape of the generated wave must be determined at a reduced voltage by connecting an oscilloscope across the forming resistor with the wire sample connected in parallel with the forming resistor. The capacitor bank is charged to the test voltage and then discharged through the forming resistor and test sample. The test sample will be considered to have passed the test if there is no distinct change in the wave shape obtained with the initial reduced voltage compared to that obtained after the application of the test voltage.
(V) The following suggested formats may be used in submitting the test results to RUS:
|
Environmental Conditioning _______ |
||||
|---|---|---|---|---|
|
FREQUENCY 1 kilohertz |
||||
|
PAIR NUMBER |
CAPACITANCE |
CONDUCTANCE |
||
|
nF/km |
(nF/mile) |
micromhos/km |
(micromhos/mile) |
|
|
Initial |
Final |
Initial |
Final |
|
|
1 |
____________ |
____________ |
____________ |
____________ |
|
2 |
____________ |
____________ |
____________ |
____________ |
|
3 |
____________ |
____________ |
____________ |
____________ |
|
Average x̄ |
____________ |
____________ |
____________ |
____________ |
Overall Percent Difference in Average x̄ Capacitance:______________ Conductance: ______________
|
Environmental Conditioning ____________ |
|||||
|---|---|---|---|---|---|
|
FREQUENCY 150 kilohertz |
|||||
|
PAIR NUMBER |
CAPACITANCE |
CONDUCTANCE |
|||
|
nF/km |
(nF/mile) |
micromhos/km |
(micromhos/mile) |
dB/km |
|
|
Initial |
Final |
Initial |
Final |
Initial |
|
|
1 |
_________ |
_________ |
_________ |
_________ |
_________ |
|
2 |
_________ |
_________ |
_________ |
_________ |
_________ |
|
3 |
_________ |
_________ |
_________ |
_________ |
_________ |
|
Average x̄ |
_________ |
_________ |
_________ |
_________ |
_________ |
Overall Percent Difference in Average x̄ Capacitance:___________ Conductance: ___________ Attenuation:___________
|
Environmental Conditioning _______________ |
||||||
|---|---|---|---|---|---|---|
|
WATER IMMERSION TEST (1 kilohertz) |
||||||
|
PAIR NUMBER |
CAPACITANCE |
CONDUCTANCE |
||||
|
nF/km |
(nF/mile) |
micromhos/km |
(micromhos/mile) |
|||
|
Initial |
24 hours |
Final |
Initial |
24 hours |
Final |
|
|
1 |
_________ |
_________ |
_________ |
_________ |
_________ |
_________ |
|
2 |
_________ |
_________ |
_________ |
_________ |
_________ |
_________ |
|
3 |
_________ |
_________ |
_________ |
_________ |
_________ |
_________ |
|
Average x̄ |
_________ |
_________ |
_________ |
_________ |
_________ |
_________ |
Overall Percent Difference in Average x̄ Capacitance:______________ Conductance: ______________
|
Water Penetration Test |
||||
|---|---|---|---|---|
|
Option A |
Option B |
|||
|
End Leakage grams |
Weight Gain grams |
End Leakage grams |
Penetration mm (in.) |
|
|
Control |
_____________ |
_____________ |
_____________ |
_____________ |
|
Heat Age |
_____________ |
_____________ |
_____________ |
_____________ |
|
Humidity Exposure |
_____________ |
_____________ |
_____________ |
_____________ |
|
Temperature Cycling |
_____________ |
_____________ |
_____________ |
_____________ |
|
Insulation Compression |
|
|---|---|
|
Failures |
|
|
Control |
________ |
|
Heat Age |
________ |
|
Humidity Exposure |
________ |
|
Temperature Cycling |
________ |
|
Jacket Slip Strength @ 50°C |
|
|---|---|
|
Load in newtons |
|
|
(pound-force) |
|
|
Control |
________ |
|
Heat Age |
________ |
|
Humidity Exposure |
________ |
|
Temperature Cycling |
________ |
|
Filler Exudation (grams) |
|
|---|---|
|
Heat Age |
________ |
|
Humidity Exposure |
________ |
|
Temperature Cycle |
________ |
|
Surge Test (kilovolts) |
|
|---|---|
|
Conductor to Conductor |
________ |
|
Shield to Conductors |
________ |
Cite this article: FindLaw.com - Code of Federal Regulations Title 7. Agriculture § 7.1755.860—Qualification Test Methods - last updated January 02, 2025 | https://codes.findlaw.com/cfr/title-7-agriculture/cfr-sect-7-1755-860-app-a/
FindLaw Codes may not reflect the most recent version of the law in your jurisdiction. Please verify the status of the code you are researching with the state legislature before relying on it for your legal needs.
A free source of state and federal court opinions, state laws, and the United States Code. For more information about the legal concepts addressed by these cases and statutes, visit FindLaw’s Learn About the Law.
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Search our directory by legal issue
Enter information in one or both fields (Required)