Compliance with MedRadio equivalent isotropic radiated power (M–EIRP) limits can be determined by measuring the radiated field strength from the transmitter type, in accordance with the rules in this section.
(a) Radiated field strength values corresponding to the M–EIRP limits in § 95.2567 are given in the table in this paragraph, for an open area test site, and for a test site equivalent to free space, such as a fully anechoic test chamber. Field strength is measured at a distance of 3 meters from the equipment under test.
|
M-EIRP limit |
Open area (mV/m) |
Free space (mV/m) |
|---|---|---|
|
1 mW |
115.1 |
57.55 |
|
25 μW |
18.2 |
9.1 |
|
250 nW |
1.8 |
0.9 |
|
100 nW |
1.2 |
0.6 |
(b) Compliance with the maximum transmitter power requirements in § 95.2567 is based on measurements using a peak detector function and measured over an interval of time when transmission is continuous and at its maximum power level. In lieu of using a peak detector function, measurement procedures that have been found to be acceptable to the FCC in accordance with § 2.947 of this chapter may be used to demonstrate compliance.
(c) For a MedRadio transmitter intended to be implanted in a human body, radiated emissions and M–EIRP measurements for transmissions by stations authorized under this section may be made in accordance with an FCC–approved human body simulator and test technique. Guidance regarding SAR measurement techniques dielectric parameters for the tissue-equivalent material can be found in the Office of Engineering and Technology (OET) Laboratory Division Knowledge Database (KDB).
(c) For a MedRadio transmitter intended to be implanted in a human body, radiated emissions and M–EIRP measurements for transmissions by stations authorized under this section may be made in accordance with an FCC–approved human body simulator and test technique. Guidance regarding dielectric parameters for the tissue-equivalent material can be found in the Office of Engineering and Technology (OET) Laboratory Division Knowledge Database (KDB).
Cite this article: FindLaw.com - Code of Federal Regulations Title 47. Telecommunication § 47.95.2569 MedRadio field strength measurements - last updated January 02, 2025 | https://codes.findlaw.com/cfr/title-47-telecommunication/cfr-sect-47-95-2569/
FindLaw Codes may not reflect the most recent version of the law in your jurisdiction. Please verify the status of the code you are researching with the state legislature before relying on it for your legal needs.
A free source of state and federal court opinions, state laws, and the United States Code. For more information about the legal concepts addressed by these cases and statutes, visit FindLaw’s Learn About the Law.
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Search our directory by legal issue
Enter information in one or both fields (Required)