(a) Applicability. Laboratories are not required to verify or establish performance specifications for any test system used by the laboratory before April 24, 2003.
(b)(1) Verification of performance specifications. Each laboratory that introduces an unmodified, FDA–cleared or approved test system must do the following before reporting patient test results:
(i) Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics:
(A) Accuracy.
(B) Precision.
(C) Reportable range of test results for the test system.
(ii) Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population.
(2) Establishment of performance specifications. Each laboratory that modifies an FDA–cleared or approved test system, or introduces a test system not subject to FDA clearance or approval (including methods developed in-house and standardized methods such as text book procedures), or uses a test system in which performance specifications are not provided by the manufacturer must, before reporting patient test results, establish for each test system the performance specifications for the following performance characteristics, as applicable:
(i) Accuracy.
(ii) Precision.
(iii) Analytical sensitivity.
(iv) Analytical specificity to include interfering substances.
(v) Reportable range of test results for the test system.
(vi) Reference intervals (normal values).
(vii) Any other performance characteristic required for test performance.
(3) Determination of calibration and control procedures. The laboratory must determine the test system's calibration procedures and control procedures based upon the performance specifications verified or established under paragraph (b)(1) or (b)(2) of this section.
(c) Documentation. The laboratory must document all activities specified in this section.
Cite this article: FindLaw.com - Code of Federal Regulations Title 42. Public Health § 42.493.1253 Standard: Establishment and verification of performance specifications - last updated January 02, 2025 | https://codes.findlaw.com/cfr/title-42-public-health/cfr-sect-42-493-1253/
FindLaw Codes may not reflect the most recent version of the law in your jurisdiction. Please verify the status of the code you are researching with the state legislature before relying on it for your legal needs.
A free source of state and federal court opinions, state laws, and the United States Code. For more information about the legal concepts addressed by these cases and statutes, visit FindLaw’s Learn About the Law.
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Search our directory by legal issue
Enter information in one or both fields (Required)