(a) Detector optimization. If necessary, follow the instrument manufacturer's instructions for initial start-up and basic operating adjustments.
(b) Calibration curve. Develop a calibration curve for each range used as follows:
(1) Zero the analyzer.
(2) Span the analyzer to give a response of approximately 90 percent of full-scale chart deflection.
(3) Recheck the zero response. If it has changed more than 0.5 percent of full scale, repeat the steps given in paragraphs (b)(1) and (b)(2) of this section.
(4) Record the response of calibration gases having nominal concentrations starting between 10 and 15 percent and increasing in at least six incremental steps to 90 percent of that range. The incremental steps are to be spaced to represent good engineering practice.
(5) Generate a calibration curve. The calibration curve shall be of fourth order or less, have five or fewer coefficients. If any range is within 2 percent of being linear a linear calibration may be used. Include zero as a data point. Compensation for known impurities in the zero gas can be made to the zero-data point. The calibration curve must fit the data points within 2 percent of point.
(6) Optional. A new calibration curve need not be generated if:
(i) A calibration curve conforming to paragraph (b)(5) of this section exists; or
(ii) The responses generated in paragraph (b)(4) of this section are within 1 percent of full scale or 2 percent of point, whichever is less, of the responses predicted by the calibration curve for the gases used in paragraph (b)(4) of this section.
(7) If multiple range analyzers are used, the lowest range used must meet the curve fit requirements below 15 percent of full scale.
Cite this article: FindLaw.com - Code of Federal Regulations Title 40. Protection of Environment § 40.89.323 NDIR analyzer calibration - last updated October 02, 2022 | https://codes.findlaw.com/cfr/title-40-protection-of-environment/cfr-sect-40-89-323/
FindLaw Codes may not reflect the most recent version of the law in your jurisdiction. Please verify the status of the code you are researching with the state legislature or via Westlaw before relying on it for your legal needs.
A free source of state and federal court opinions, state laws, and the United States Code. For more information about the legal concepts addressed by these cases and statutes, visit FindLaw's Learn About the Law.
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Search our directory by legal issue
Enter information in one or both fields (Required)