Learn About The Law
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Current as of January 02, 2025 | Updated by Findlaw Staff
|
Subpart E procedure |
Performance test |
Performance specification |
Test conditions |
Part 50, appendix L |
|---|---|---|---|---|
|
reference |
||||
|
§ 53.52 Sample leak check test |
Sampler leak check facility |
External leakage: 80 mL/min, max |
Controlled leak flow rate of 80 mL/ min |
Sec. 7.4.6. |
|
Internal leakage: 80 mL/min, max |
||||
|
§ 53.53 Base flow rate test |
Sample flow rate |
1. 16.67 ±5%, L/min |
(a) 6-hour normal operational test plus flow rate cut-off test |
Sec. 7.4.1. |
|
1. Mean |
2. 2%, max |
(b) Normal conditions |
Sec. 7.4.2. |
|
|
2. Regulation |
3. 2%, max |
(c) Additional 55 mm Hg pressure drop to simulate loaded filter |
Sec. 7.4.3. |
|
|
3. Meas accuracy |
4. 0.3% max |
(d) Variable flow restriction used for cut-off test |
Sec. 7.4.4. |
|
|
4. CV accuracy |
5. Flow rate cut-off if flow rate deviates more than 10% from design flow rate for >60 ±30 seconds |
Sec. 7.4.5. |
||
|
5. Cut-off |
||||
|
§ 53.54 Power interruption test |
Sample flow rate |
1. 16.67 ±5%, L/min |
(a) 6-hour normal operational test |
Sec. 7.4.1. |
|
1. Mean |
2. 2%, max |
(b) Nominal conditions |
Sec. 7.4.2. |
|
|
2. Regulation |
3. 2%, max |
(c) Additional 55 mm Hg pressure drop to simulate loaded filter |
Sec. 7.4.3. |
|
|
3. Meas. accuracy |
4. 0.3% max |
(d) 6 power interruptions of various durations |
Sec. 7.4.5. |
|
|
4. CV accuracy |
5. ±2 min if >60 seconds. |
Sec. 7.4.12. |
||
|
5. Occurrence time of power interruptions |
6. ±20 seconds |
Sec. 7.4.13. |
||
|
6. Elapsed sample time |
7. ±2%, max |
Sec. 7.4.15.4. |
||
|
7. Sample volume |
Sec. 7.4.15.5. |
|||
|
§ 53.55 Temperature and line voltage test |
Sample flow rate |
1. 16.67 ±5%, L/min |
(a) 6-hour normal operational test |
Sec. 7.4.1. |
|
1. Mean |
2. 2%, max |
(b) Normal conditions |
Sec. 7.4.2. |
|
|
2. Regulation |
3. 2%, max |
(c) Additional 55 mm Hg pressure drop to simulate loaded filter |
Sec. 7.4.3. |
|
|
3. Meas. accuracy |
4. 0.3% max |
(d) Ambient temperature at -20 and +40 °C |
Sec. 7.4.5. |
|
|
4. CV accuracy |
5. 2 °C |
(e) Line voltage: 105 Vac to 125 Vac |
Sec. 7.4.8. |
|
|
5. Temperature meas. accuracy |
Sec. 7.4.15.1. |
|||
|
6. Proper operation. |
||||
|
§ 53.56 Barometric pressure effect test |
Sample flow rate |
1. 16.67 ±5%, L/min |
(a) 6-hour normal operational test |
Sec. 7.4.1. |
|
1. Mean |
2. 2%, max |
(b) Normal conditions |
Sec. 7.4.2. |
|
|
2. Regulation |
3. 2%, max |
(c) Additional 55 mm Hg pressure drop to simulate loaded filter |
Sec. 7.4.3. |
|
|
3. Meas. accuracy |
4. 0.3% max |
(d) Barometric pressure at 600 and 800 mm Hg |
Sec. 7.4.5. |
|
|
4. CV accuracy |
5. 10 mm Hg |
Sec. 7.4.9. |
||
|
5. Pressure meas. accuracy |
||||
|
6. Proper operation. |
||||
|
§ 53.57 Filter temperature control test |
1. Filter temp. meas. accuracy |
1. 2 °C |
(a) 4-hour simulated solar radiation, sampling |
Sec. 7.4.8. |
|
2. Ambient temp. meas. accuracy |
2. 2 °C |
(b) 4-hour simulated solar radiation, non-sampling |
Sec. 7.4.10. |
|
|
3. Filter temp. control accuracy, sampling and non-sampling |
3. Not more than 5 °C above ambient temp. for more than 30 min |
(c) Solar flux of 1000 ±50 W/m2 |
Sec. 7.4.11. |
|
|
§ 53.58 Field precision test |
1. Measurement precision |
1. Pj < 2 μg/m3 or RPj < 5% |
(a) 3 collocated samplers at 1 site for at least 10 days |
Sec. 5.1. |
|
2. Storage deposition test for sequential samplers |
2. 50 μg max. average weight gain/blank filter |
(b) PM2.5 conc. > 3 μg/m3 |
Sec. 7.3.5. |
|
|
(c) 24– or 48–hour samples |
Sec. 8. |
|||
|
(d) 5– or 10–day storage period for inactive stored filters |
Sec. 9. |
|||
|
Sec. 10. |
||||
|
The Following Requirement Is Applicable to Class I Candidate Equivalent Methods Only |
||||
|
§ 53.59 Aerosol transport test |
Aerosol transport |
97%, min. for all channels. |
Determine aerosol transport through any new or modified components with respect to the reference method sampler before the filter for each channel |
Cite this article: FindLaw.com - Code of Federal Regulations Title 40. Protection of Environment 40 CFR Pt. 53, Subpt. E, Tbl. E-1 Table E-1 to Subpart E of Part 53—Summary of Test Requirements for Reference and Class I Equivalent Methods for PM 2.5 and PM 10-2.5 - last updated January 02, 2025 | https://codes.findlaw.com/cfr/title-40-protection-of-environment/cfr-pt-40-53-subpt-e-tbl-e-1/
FindLaw Codes may not reflect the most recent version of the law in your jurisdiction. Please verify the status of the code you are researching with the state legislature before relying on it for your legal needs.
A free source of state and federal court opinions, state laws, and the United States Code. For more information about the legal concepts addressed by these cases and statutes, visit FindLaw’s Learn About the Law.
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Search our directory by legal issue
Enter information in one or both fields (Required)