(a) Identification. An image processing device for estimation of external blood loss is a device to be used as an aid in estimation of patient external blood loss. The device may include software and/or hardware that is used to process images capturing externally lost blood to estimate the hemoglobin mass and/or the blood volume present in the images.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. Demonstration of the performance characteristics must include a comparison to a scientifically valid alternative method for measuring deposited hemoglobin mass. The following use conditions must be tested:
(i) Lighting conditions;
(ii) Range of expected hemoglobin concentrations;
(iii) Range of expected blood volume absorption; and
(iv) Presence of other non-sanguineous fluids (e.g., saline irrigation fluid).
(2) Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by intended users of the device.
(3) Appropriate analysis and non-clinical testing must validate the electromagnetic compatibility (EMC) and wireless performance of the device.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Software display must include an estimate of the cumulative error associated with estimated blood loss values.
(6) Labeling must include:
(i) Warnings, cautions, and limitations needed for safe use of the device;
(ii) A detailed summary of the performance testing pertinent to use of the device, including a description of the bias and variance the device exhibited during testing;
(iii) The validated surgical materials, range of hemoglobin mass, software, hardware, and accessories that the device is intended to be used with; and
(iv) EMC and wireless technology instructions and information.
Cite this article: FindLaw.com - Code of Federal Regulations Title 21. Food and Drugs § 21.880.2750 Image processing device for estimation of external blood loss - last updated January 02, 2025 | https://codes.findlaw.com/cfr/title-21-food-and-drugs/cfr-sect-21-880-2750/
FindLaw Codes may not reflect the most recent version of the law in your jurisdiction. Please verify the status of the code you are researching with the state legislature before relying on it for your legal needs.
A free source of state and federal court opinions, state laws, and the United States Code. For more information about the legal concepts addressed by these cases and statutes, visit FindLaw’s Learn About the Law.
Get help with your legal needs
FindLaw’s Learn About the Law features thousands of informational articles to help you understand your options. And if you’re ready to hire an attorney, find one in your area who can help.
Search our directory by legal issue
Enter information in one or both fields (Required)